Accurate mass flow, guaranteed process integrity
The SumoFlo Coriolis Flow Meter from CPC Biotech establishes an uncompromising standard for critical fluid management in biomanufacturing. As a uniquely gamma-sterilizable Coriolis sensor, it is designed for seamless integration into single-use workflows, providing foundational sterility assurance across sensitive processes. Utilizing advanced Coriolis technology, it delivers true mass flow measurement with inherent accuracy that is fundamentally independent of variations in fluid properties, such as density or viscosity. This capability eliminates the need for recalibration, which simplifies validation and ensures precise, reliable process control, enhancing operational flexibility and drastically reducing the risk of cross-contamination.
Sterility assurance
- Gamma-sterilizable Coriolis sensor
- Disposable, drop-in design for single-use workflows
- USP Class VI PEEK wetted materials (also meets 661, 788)
- Reduces cross-contamination risk and minimizes cleaning validation
Measurement performance
- Best-in-class ±1% mass flow accuracy
- Independent of fluid properties (density, viscosity)
- Unaffected by flow regime or velocity profile
- Accurate measurement of bubbly or frothy media
Application flexibility
- Wide flow range: 20 g/min to 100 kg/min
- Multiple sensor sizes: 1/8" to 1"
- Process connection options: barb or tri-clamp
- Optional display transmitter for on-site monitoring (flow, density, temp)
Technical data
| Model | 8100 - 082 | 8100 - 152 |
| Accuracy | ±1% of rate for 10% to 100% of full scale rated flow rate ±(1% of rate + Z.O.S) for < 10% of full scale rated flow rate |
|
| Temperature | Ambient: 0-50 °C Fluid: 4-40 °C |
|
| Operating pressure | 80 psig (max.) | 80 psig (max.) |
| Flow range | 2,000 - 20,000 g/min | 7,500 - 75,000 g/min |
| Zero offset stability (Z.O.S.) | 5 g/min | 18.75 g/min |